Guide · Standards

    IEC 62443 Evidence Table for Medical Device Manufacturers

    Map IEC 62443-4-1 practices and 62443-4-2 foundational requirements to controlled medical device evidence, owners, and review questions.

    Hero illustration for the Standards article: IEC 62443 Evidence Table for Medical Device Manufacturers
    On this page
    Christian Espinosa, Founder & CEO at Blue Goat Cyber

    By Christian Espinosa, MBA

    Founder & CEO · Blue Goat Cyber

    Published:

    Key Takeaways

    • IEC 62443 is an industrial automation security series; medical device teams mostly use part 4-1 for development practices and part 4-2 for component technical requirements.
    • Part 4-1 organizes secure development into eight practices; part 4-2 organizes technical requirements under seven foundational requirements.
    • Security levels in part 4-2 describe capability against attacker types, not a regulatory pass mark.
    • IEC 81001-5-1 is the health software lifecycle standard; IEC 62443 evidence can support it but is not a substitute for device risk and harm analysis.
    • Check the FDA recognized consensus standards database for the current recognition status of each part before citing it.
    Direct Answer

    What evidence does IEC 62443 call for on a medical device?

    For a medical device, IEC 62443 evidence usually comes from two parts. IEC 62443-4-1 asks for proof of a secure development process across eight practices, and IEC 62443-4-2 asks for proof that a component meets technical requirements under seven foundational requirements. Use the tables below to connect each area to controlled records, an owner, and the question a reviewer will ask. Pair it with the IEC 81001-5-1 evidence table, which is the health software lifecycle standard.

    This page is an evidence planning aid, not a clause-by-clause conformity checklist. Clause wording, requirement enhancements, and security level details are in the purchased standards.

    What this table does and does not establish

    • It shows the kinds of records that typically demonstrate each area.
    • It does not claim conformity, certification, or FDA acceptance.
    • It does not replace security risk assessment under ANSI/AAMI SW96 or the link to harm under ISO 14971.
    • The manufacturer decides which parts apply, owns the records, and approves risk decisions.

    Which IEC 62443 parts matter for devices

    PartScopeTypical device use
    62443-4-1Secure product development lifecycle requirements for product suppliersDevelopment process evidence, often alongside IEC 81001-5-1
    62443-4-2Technical security requirements for components (embedded devices, host devices, network devices, and software applications)Component capability evidence, such as authentication, logging, and integrity controls
    62443-3-3System security requirements and security levelsSystems made of several components, such as a device plus gateway and hospital integration
    62443-3-2Security risk assessment for system design, including zones and conduitsSegmenting a multi-component system and assigning target security levels
    62443-2-4Security program requirements for service providersIntegrators or service organizations supporting deployed systems

    IEC 62443-4-1 practice to evidence table

    PracticeExample controlled evidenceOwner and review question
    1. Security management (SM)Security development plan, roles, competency records, process scope, and third-party component controlsEngineering / QA: Is the process defined, staffed, and applied to this product?
    2. Specification of security requirements (SR)Product security context, threat model, security requirements, and requirement reviewsProduct security: Are requirements derived from the intended environment and threats?
    3. Secure by design (SD)Architecture, trust boundaries, defense in depth rationale, and design reviewsSystem engineering: Does the design show where each control sits and why?
    4. Secure implementation (SI)Secure coding standard, code review records, static analysis results, and dependency inventorySoftware engineering: Are coding rules applied and exceptions recorded?
    5. Security testing (SVV)Requirement tests, threat mitigation tests, vulnerability scanning, fuzzing where relevant, penetration testing, and tester independence recordsProduct security: Does test evidence cover the released build and its limitations?
    6. Management of security-related issues (DM)Intake process, triage records, severity assessment, disposition, and closure evidencePSIRT: Can every reported issue be traced from receipt to closure?
    7. Security update management (SUM)Update qualification, delivery integrity, timeliness commitments, and customer notification recordsMaintenance engineering: Are updates tested, signed, and communicated?
    8. Security guidelines (SG)Hardening guide, secure configuration, defense in depth expectations for the user, and decommissioning guidanceLabeling / product security: Do users get what they need to deploy and retire the product securely?

    Practice 5 is the testing practice. Blue Goat Cyber performs security and penetration testing for it; general software verification and validation remains with the manufacturer.

    IEC 62443-4-2 foundational requirement to evidence table

    Foundational requirementWhat it coversExample device evidence
    FR 1: Identification and authentication control (IAC)Who or what can connectAccount model, credential storage design, authentication tests, and default credential removal
    FR 2: Use control (UC)What an authenticated user can doRole and permission matrix, authorization tests, and session controls
    FR 3: System integrity (SI)Protection against unauthorized changesSecure boot, code signing, update integrity checks, and tamper test results
    FR 4: Data confidentiality (DC)Protection of data in transit and at restEncryption design, key management, and protocol test results
    FR 5: Restricted data flow (RDF)Segmentation and limiting connectionsPort and service inventory, interface restrictions, and network segmentation guidance
    FR 6: Timely response to events (TRE)Logging and auditSecurity event list, log protection, and log export tests
    FR 7: Resource availability (RA)Resilience against denial of serviceRate limiting, resource exhaustion tests, and safe degraded behavior

    Security levels in plain terms

    Part 4-2 describes capability security levels from SL 1 (protection against casual or coincidental violation) to SL 4 (protection against sophisticated attackers with extended resources). A target level comes from the system risk assessment. It is not an FDA pass mark, and claiming a level requires evidence for each applicable requirement and enhancement.

    How IEC 62443 and IEC 81001-5-1 fit together

    TopicIEC 62443-4-1 / 4-2IEC 81001-5-1
    OriginIndustrial automation and control systemsHealth software and health IT systems
    LifecycleEight development practicesDevelopment, maintenance, risk, configuration, and problem resolution processes aligned with IEC 62304
    Technical controlsDetailed in 4-2 by foundational requirementNot specified as a component requirement catalog
    Patient harmNot addressedConnects to ISO 14971 where security affects safety

    IEC 81001-5-1 draws on the structure of 62443-4-1, so one set of records can often serve both if each record states which requirement it supports. Assess each standard separately.

    Worked example: one control traced across both standards

    LinkExample recordIEC 62443 referenceIEC 81001-5-1 connection
    ThreatTM-07: a forged firmware image is installed through the service port4-1 SR, SDThreat and architecture records
    RequirementSEC-31: the bootloader rejects unsigned images4-2 FR 3Security requirement
    DesignDES-31: signature check with protected root key4-1 SD; 4-2 FR 3Architecture and design
    TestTEST-31: altered and unsigned images are rejected on the release build4-1 SVVSecurity verification evidence
    IssueFIND-04: debug interface left enabled on production units4-1 DMProblem resolution
    Update and guidanceREL-12 closes the interface; hardening guide updated4-1 SUM, SGMaintenance and release

    FDA context

    The FDA's February 3, 2026 premarket cybersecurity guidance, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, does not require IEC 62443 conformance. Manufacturers may use recognized consensus standards to support a submission. Recognition status changes, so check the FDA recognized consensus standards database for each part and edition before citing it in an eSTAR.

    Practical gap review

    1. List which parts and editions you intend to cite.
    2. For each 4-1 practice, name the procedure and the product record that shows it was followed.
    3. For each 4-2 foundational requirement, record whether it applies, the control, and the test.
    4. Confirm test evidence matches the release build.
    5. Link security findings to SW96 risk records and, where relevant, ISO 14971 hazards.
    6. Keep one evidence index that states which standard each record supports.

    How Blue Goat Cyber helps

    We build threat models, security risk assessments, and penetration test evidence that map to IEC 62443 and IEC 81001-5-1 for FDA premarket submissions. We do not certify conformity. See medical device threat modeling and penetration testing.

    FAQ

    Do medical devices need IEC 62443?

    No law or FDA guidance requires it. Many manufacturers use 62443-4-1 and 4-2 because they give detailed development practices and technical requirements that support IEC 81001-5-1 and FDA submissions.

    Is IEC 62443 the same as IEC 81001-5-1?

    No. IEC 81001-5-1 is written for health software and connects to medical device risk management. IEC 62443 comes from industrial control systems. They share structure, and records can serve both.

    What security level should a medical device target?

    The target comes from your system risk assessment, the intended use environment, and the attackers you expect. No security level is mandated by the FDA.

    Can one test report cover both standards?

    Yes, if the report identifies the build, the requirements tested, the results, and the limitations, and your evidence index maps it to each standard.

    Sources and primary references

    Sources & references

    Primary sources cited in this article. Links open in a new tab.

    1. FDA recognized consensus standards database- U.S. FDA
    2. Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions- U.S. FDA
    Related. EU Medical Device Cybersecurity Guide