MEDICAL DEVICE REFERENCE
FDA guidance comparison
Medical-device cybersecurity, software, AI, and change-control documents: who they apply to, the evidence they recommend, and what is actually required by law.
Reviewed . Selected relevant documents, not the entire FDA guidance library. Device-specific applicability and regulatory decisions remain with the manufacturer.
Law and regulations
Binding when applicable. Section 524B, submission rules, and reporting regulations must be assessed separately from guidance recommendations.
Final guidance
The FDA's current recommendations. Nonbinding unless a cited statute or regulation creates the obligation; an appropriate alternative approach may be used.
Draft guidance
Proposals for comment, not for implementation. Do not treat draft recommendations as final requirements.
Section 524B: the legal baseline
For covered cyber-device marketing submissions, including 510(k), PMA, PDP, De Novo, and HDE, Section 524B requires the following. Relevant supplements and Special or Abbreviated 510(k)s are also covered. IDE submissions are not covered by Section 524B; investigational-device cybersecurity is addressed separately in Appendix 3 of the 2026 guidance.
- Submit a plan to monitor, identify, and address postmarket cybersecurity vulnerabilities and exploits, including coordinated vulnerability disclosure and related procedures.
- Design, develop, and maintain processes and procedures providing reasonable assurance that the device and related systems are cybersecure, and make updates and patches available on the statutory risk-based schedule.
- Provide an SBOM including commercial, open-source, and off-the-shelf software components.
The statute distinguishes known unacceptable vulnerabilities, addressed on a reasonably justified regular cycle, from critical vulnerabilities that could cause uncontrolled risk, addressed as soon as possible out of cycle. It does not set one universal patch deadline. Section 524B(b)(4) also covers other requirements the Secretary may require through regulation.
FDA cybersecurity and Section 524B informationCompare guidance documents
7 documents
| Document and scope | Key recommendations | Evidence and boundaries |
|---|---|---|
Final guidance · PremarketCybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket SubmissionsDevices with cybersecurity considerations; Section 524B duties apply specifically to covered cyber-device marketing submissions. Appendix 3 addresses investigational devices. FDA source |
| Evidence to prepare Threat model, architecture views, security risk file, SBOM, test results, labeling, and postmarket plan, scoped to the device and submission. Important boundary Supersedes the June 27, 2025 version. Recommendations are nonbinding; Section 524B obligations are law. IDEs are not covered marketing submissions under Section 524B. |
Final guidance · PostmarketPostmarket Management of Cybersecurity in Medical DevicesCybersecurity management for medical devices already distributed, including legacy devices. FDA source |
| Evidence to prepare Monitoring records, vulnerability assessments, disclosure records, remediation verification, user communications, and reporting assessments. Important boundary Not replaced by the 2026 premarket guidance. Conditional 30/60-day enforcement-discretion criteria are not universal patch deadlines or automatic reporting exemptions. |
Final guidance · AI and changesMarketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software FunctionsManufacturers proposing an optional PCCP for AI-enabled device software functions in a marketing submission. FDA source |
| Evidence to prepare Proposed modifications, modification protocol, impact assessment, acceptance criteria, and implementation records. Important boundary PCCPs are optional and need FDA authorization. Changes outside a plan need a device-specific change assessment, not an automatic new-submission conclusion. |
Draft guidance · AI and changesArtificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission RecommendationsProposed lifecycle and submission recommendations for AI-enabled device software functions. FDA source |
| Evidence to prepare Proposed planning evidence: data descriptions, performance evaluations, risk assessment, user information, and monitoring plans. Important boundary Draft, not for implementation. It is not a final mandatory checklist and does not require every AI device to have a PCCP. |
Draft guidance · AI and changesPredetermined Change Control Plans for Medical DevicesProposed broader PCCP approach for medical devices, beyond the final AI-specific guidance. FDA source |
| Evidence to prepare Proposed plan boundaries, modification methods, acceptance criteria, impact assessment, and change records. Important boundary Draft, not for implementation. Do not apply it as final guidance for every non-AI software or cybersecurity update; a PCCP remains optional. |
Final guidance · PremarketContent of Premarket Submissions for Device Software FunctionsSoftware documentation in marketing submissions for devices with software functions, whether or not they are cyber devices. FDA source |
| Evidence to prepare Software documentation-level rationale, requirements, architecture, risk records, testing documentation, version history, and anomaly list. Important boundary Software documentation and cybersecurity evidence describe the same device but serve different review questions. This is not a claim that Blue Goat Cyber provides general software validation services. |
Final guidance · AI and changesDeciding When to Submit a 510(k) for a Software Change to an Existing DeviceSoftware changes to an existing device evaluated under the 510(k) change framework; not a substitute for PMA-specific change rules. FDA source |
| Evidence to prepare Change description, risk assessment, verification and validation evidence, and documented submission decision. Important boundary A cybersecurity patch does not automatically require, or automatically avoid, a new 510(k). The manufacturer owns the device-specific decision. |