Blue Goat CyberBlue Goat CyberSMMedical Device Cybersecurity
    K
    Goat Feed · Guidance

    Flow Neuroscience FL-100

    Successful exploitation of this vulnerability could allow an attacker within Bluetooth range to manipulate brain stimulation parameters and override safety limits.

    Back to the daily feed
    Guidance
    notable

    View CSAF

    Summary

    Successful exploitation of this vulnerability could allow an attacker within Bluetooth range to manipulate brain stimulation parameters and override safety limits.

    The following versions of Flow Neuroscience FL-100 are affected:

    • Flow Neuroscience FL-100
    • Halo Neuroscience FL-100
    CVSS Vendor Equipment Vulnerabilities
    v3 8.1 Flow Neuroscience Flow Neuroscience FL-100 Use of Hard-coded Credentials

    Background

    • Critical Infrastructure Sectors: Healthcare and Public Health
    • Countries/Areas Deployed: Worldwide
    • Company Headquarters Location: Sweden

    Vulnerabilities

    CVE-2026-18164

    An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarilymanipulate brain stimulation parameters and state.

    View CVE Details


    Affected Products

    Flow Neuroscience FL-100
    Vendor:
    Flow Neuroscience
    Product Version:
    Flow Neuroscience Flow Neuroscience FL-100:
    Product Status:
    known_affected
    Remediations

    Mitigation
    Users are encouraged to install the latest firmware updates provided by Flow Neuroscience via the Flow app.

    Relevant CWE: CWE-798 Use of Hard-coded Credentials


    Metrics

    CVSS Version Base Score Base Severity Vector String
    3.1 8.1 HIGH CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H
    4.0 7.2 HIGH CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N

    Acknowledgments

    • A.C. Buglione reported this vulnerability to CISA

    Legal Notice and Terms of Use

    This product is provided subject to this Notification (https://www.cisa.gov/notification) and this Privacy & Use policy (https://www.cisa.gov/privacy-policy).


    Recommended Practices

    CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability.

    Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.

    Locate control system networks and remote devices behind firewalls and isolating them from business networks.

    When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

    CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

    CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

    CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

    Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

    Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

    CISA also recommends users take the following measures to protect themselves from social engineering attacks:

    Do not click web links or open attachments in unsolicited email messages.

    Refer to Recognizing and Avoiding Email Scams for more information on avoiding email scams.

    Refer to Avoiding Social Engineering and Phishing Attacks for more information on social engineering attacks.

    No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.


    Revision History

    • Initial Release Date: 2026-08-13
    Date Revision Summary
    2026-08-13 1 Initial Publication

    Legal Notice and Terms of Use

    Citations
    #icsma#advisory
    Need help acting on this?

    Talk to Blue Goat Cyber

    Penetration testing, SBOM, threat modeling, and 524B submission support for MedTech.