Goat Feed · Guidance

    J4Care dcm4che

    Successful exploitation of these vulnerabilities could allow an attacker to delete stored medical images, disclose patient information, inject forged studies, reassign patient records, overwrite files, and cause a denial-of-service condition.

    Back to the daily feed
    Guidance
    notable

    Blue Goat analysis

    Scored with Blue Goat's medical device cybersecurity risk methodology: STRIDE classification and CVSS v4.0 exploitability, with a severity-of-harm readout for your own ISO 14971 hazard analysis.

    The dcm4chee-arc-light archive is an open-source PACS used to store and serve medical images. CISA's advisory covers six vulnerabilities in version 5.35.1. Two of them go beyond data exposure: they let an unauthenticated attacker change which patient a study belongs to.

    CVEWhat an attacker can doSTRIDECVSS v4.0 ExploitabilitySecurity impact
    CVE-2026-105667In the default configuration, the HL7 listener accepts patient-merge messages (ADT^A40) without authenticating the sender, so a remote sender can reassign studies and related workflow records (CWE-306)Spoofing, TamperingHighHigh
    CVE-2026-105279In the default configuration, the archive accepts DICOM associations from unregistered callers, who can retrieve patient information and images, and file fabricated study content under an existing patient identity (CWE-306)Spoofing, Tampering, Information disclosureHighHigh
    CVE-2026-97851A storage REST service lets locations still holding DICOM instances be moved to deletion states. The purge scheduler then deletes the stored object while the database still lists the instance (CWE-862)Tampering, Denial of serviceHighHigh
    CVE-2026-107621A crafted or truncated JPEG stream can put the parser into an infinite loop (CWE-835)Denial of serviceHighHigh
    CVE-2026-105273A malformed MP4 box can stall parsing indefinitely (CWE-835)Denial of serviceHighHigh
    CVE-2026-101021Crafted multipart input forces excessive processing (CWE-407)Denial of serviceHighHigh

    All six published vectors are network-reachable with no privileges and no user action, so Exploitability reads High. Each has at least one High impact. On Blue Goat's security matrix, all six are Critical as published.

    The published vectors describe the software, not your device. If the service is reachable only on a segmented clinical network or behind authentication, score your own deployment with the CVSS v4.0 environmental metrics before you rate your risk.

    Severity from your hazard analysisSafety risk at High exploitability
    NegligibleMedium
    Minor or SeriousHigh
    Critical or CatastrophicVery High

    The two default-configuration flaws matter most clinically. A study filed under the wrong patient, or fabricated content attached to a real one, puts incorrect data in front of a clinician. The purge flaw can leave the archive listing images that no longer exist.

    What to check now

    1. Search your SBOM for dcm4chee-arc-light 5.35.1, or any product built on it.
    2. Check your configuration against the two default-configuration flaws: whether unregistered calling AE titles are accepted, and whether the HL7 listener authenticates senders.
    3. Update to release 5.35.2 or later, which J4Care recommends.
    4. If you're not affected, record that in a VEX statement with the justification.
    5. Re-score each affected threat, and carry patient-harm results into your ISO 14971 file.

    This analysis is informational. It is based on CISA's published advisory and does not assess any specific manufacturer's device.

    Other items reviewed by the Blue Goat team.

    Need help acting on this?

    Talk to Blue Goat Cyber

    Penetration testing, SBOM, threat modeling, and 524B submission support for MedTech.