    Medixant RadiAnt DICOM

    Successful exploitation of this vulnerability could allow an attacker to cause the application to crash if a maliciously crafted DICOM file is opened.

    Back to the daily feed
    Guidance
    notable

    Blue Goat analysis

    Scored with Blue Goat's medical device cybersecurity risk methodology: STRIDE classification and CVSS v4.0 exploitability, with a severity-of-harm readout for your own ISO 14971 hazard analysis.

    RadiAnt is a DICOM viewer used to open and review medical images. CISA's advisory covers one vulnerability in versions below 2025.2.

    CVEWhat an attacker can doSTRIDECVSS v4.0 ExploitabilitySecurity impact
    CVE-2026-17264Opening a crafted DICOM file with malicious JPEG-compressed pixel data triggers a heap out-of-bounds write, which may allow remote code execution (CWE-787)Tampering, Elevation of privilegeMediumMedium

    The published vector is network-reachable with no privileges, but a user has to open the file. That's why Exploitability reads Medium. CISA's published impact is Low availability only, so security impact reads Medium, and the security matrix gives Medium as published.

    The description is more serious than the published impact. CISA's own text says the flaw may allow remote code execution, but the vector scores no confidentiality or integrity impact. If your product parses DICOM pixel data, don't inherit this vector. Score the impact for your own code path.

    Severity from your hazard analysisSafety risk at Medium exploitability
    NegligibleLow
    Minor or SeriousMedium
    Critical or CatastrophicHigh

    Any software that parses medical images from outside sources faces the same risk: the image file itself becomes the attack.

    What to check now

    1. Check whether your product uses RadiAnt below 2025.2, or a JPEG decoder in its DICOM pipeline with the same pattern.
    2. Fuzz your DICOM and image parsing paths, especially compressed pixel data.
    3. Update to version 2026.1, which CISA recommends, and open DICOM files only from trusted sources.
    4. If you're not affected, record that in a VEX statement with the justification.
    5. Re-score each affected threat in your cybersecurity risk assessment, and carry any patient-harm result into your ISO 14971 file.

    This analysis is informational. It is based on CISA's published advisory and does not assess any specific manufacturer's device.