Goat Feed · Guidance

    OHIF Viewers DICOM

    Successful exploitation of this vulnerability in a custom integration version could allow an attacker to steal an authenticated clinician's token via a crafted link.

    Back to the daily feed
    Guidance
    notable

    Blue Goat analysis

    Scored with Blue Goat's medical device cybersecurity risk methodology: STRIDE classification and CVSS v4.0 exploitability, with a severity-of-harm readout for your own ISO 14971 hazard analysis.

    OHIF is an open-source, web-based medical image viewer framework that many imaging products and SaMD front ends build on. CISA's advisory covers one vulnerability in versions below 3.12.0.

    CVEWhat an attacker can doSTRIDECVSS v4.0 ExploitabilitySecurity impact
    CVE-2026-12473Two data sources in the default configuration (DICOMWebProxy and DICOMJSON) fetch an arbitrary URL without validation, and OHIF attaches the logged-in user's OIDC bearer token to that request, sending it to an attacker's server (CWE-918)Information disclosure, SpoofingMediumHigh

    The published vector is network-reachable with no privileges, but it requires a user to take an action. So Exploitability reads Medium. Confidentiality impact is High on both the viewer and the systems the token unlocks. On Blue Goat's security matrix, that's High as published.

    The published vectors describe the software, not your device. If the service is reachable only on a segmented clinical network or behind authentication, score your own deployment with the CVSS v4.0 environmental metrics before you rate your risk.

    Severity from your hazard analysisSafety risk at Medium exploitability
    NegligibleLow
    Minor or SeriousMedium
    Critical or CatastrophicHigh

    A stolen token lets an attacker act as the user whose session leaked it, with whatever access that user has to images and patient data.

    What to check now

    1. Search your SBOM for OHIF Viewer below 3.12.0.
    2. Check whether your deployment enables the DICOMWebProxy or DICOMJSON data sources. CISA notes that DICOMweb data sources aren't affected.
    3. Update to 3.12.2 or later. CISA reports the maintainer released the fix in 3.12.2.
    4. If you're not affected, record that in a VEX statement with the justification.
    5. Re-score the threat, and carry any patient-harm result into your ISO 14971 file.

    This analysis is informational. It is based on CISA's published advisory and does not assess any specific manufacturer's device.

    Other items reviewed by the Blue Goat team.

    Need help acting on this?

    Talk to Blue Goat Cyber

    Penetration testing, SBOM, threat modeling, and 524B submission support for MedTech.