    FDA Recall: Boston Scientific Corporation, System, Image Processing, Radiological

    Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that…

    Back to the daily feed
    Software Quality
    infoBoston Scientific

    Recalling firm: Boston Scientific Corporation

    Device class: 2

    Device name: System, Image Processing, Radiological

    Event initiated: 2026-06-19

    FDA posted: 2026-07-22

    Recall ID: Z-2771-2026

    Product: Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

    Reason for recall

    Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.