Software Quality
infoGoat Feed · Software Quality
FDA Recall: Abiomed, Inc., Temporary Non-Roller Type Left Heart Support Blood Pump
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to…