Before we test, we check that your threat model, hazard analysis, and eSTAR architecture diagram line up, so every finding traces from diagram to threat to patient harm.
In the first week of a medical device penetration test, we review your threat model, hazard analysis, security architecture views, and eSTAR software architecture diagram against the checklist below. You get a short review memo with each item rated, in three parts.
Met
Ready for testing and traceable.
Partly met
Testing can proceed; fix before you submit.
Not met
A traceability gap a reviewer is likely to question.
Is the threat model complete enough to aim the test at the right places?
| ID | What we check | Why it matters for the test |
|---|---|---|
| A1 | The system is modeled with data flows, data stores, processes, external entities, and trust boundaries. | Without the full system, test targets are guesses. |
| A2 | Every entry point in scope is represented, with protocols named on high-value flows. | Each entry point becomes a test target. |
| A3 | Threats are listed per element using STRIDE, each with a stable threat ID. | Findings in the report trace back to these IDs. |
| A4 | STRIDE sorts threats into types; it is not used to score them. | Mixing the two hides which threats matter most. |
| A5 | Risk is scored on exploitability, not likelihood or probability. | The FDA expects exploitability for security risk. |
| A6 | Each threat has a control strategy: eliminate, mitigate, accept, or transfer. | Mitigated threats get tests that prove the control works. |
| A7 | Assumptions and security context are stated, including the update path and system boundaries. | Unstated assumptions become untested gaps. |
Does each security threat connect to patient harm?
| ID | What we check | Why it matters for the test |
|---|---|---|
| B1 | Cybersecurity-initiated hazards are included, especially the patient effects of losing confidentiality, integrity, or availability. | Test findings are rated by patient harm, not CVSS alone. |
| B2 | Severity is rated on the ISO 14971 scale, separately from security scoring. | Severity and exploitability are two different questions. |
| B3 | Threat IDs map to hazard IDs. | This is the link a reviewer follows from finding to harm. |
Do the eSTAR software architecture diagram, the security architecture views, and the threat model describe the same system?
| ID | What we check | Why it matters for the test |
|---|---|---|
| C1 | The security views start from the same system as the eSTAR software architecture diagram. | Two different pictures of one device draw reviewer questions. |
| C2 | Components use the same names in the diagram, views, threat model, and SBOM. | Mismatched names break traceability. |
| C3 | Every external interface in the diagram crosses a trust boundary in the views and appears in the threat model. | Missing interfaces are missing test targets. |
| C4 | The global system view shows which paths could reach many devices from one entry point. | Multi-patient harm paths get their own tests. |
| C5 | The updatability view matches the update path in the threat model. | Update-path tests, such as unsigned or older firmware, come from here. |
| C6 | The diagram, views, threat model, and SBOM all describe the same software version. | The test must run on the version you submit. |
The eSTAR asks for a software architecture diagram in the software section and security architecture views in the cybersecurity section. They should be two views of one system. Part C checks that rule before testing, because the test targets come from the interfaces and trust boundaries in those pictures. See FDA cybersecurity architecture views for the side-by-side comparison, and eSTAR software documents and cybersecurity for how the two sections stay consistent.
Try the free Threat Model Starter, read the cybersecurity hazard analysis guide, or book a threat model gap assessment for a deeper review.