Premarket · Infusion / Drug Delivery

    Threat Modeling for Infusion & Drug Delivery

    Threat models for smart pumps, syringe drivers, and connected drug-delivery systems - DERS libraries, EHR auto-program, and cloud telemetry covered.

    Aligned to FDA's February 3, 2026 premarket cybersecurity guidance.

    How this applies to Infusion / Drug Delivery

    Infusion-pump threat modeling has to take seriously a fact other segments don't: the highest-impact cyber events in this segment have come not from device-level RCEs, but from drug-error reduction software (DERS) library tampering, EHR auto-programming integrity failures, and cloud-side configuration changes that propagated to thousands of fielded pumps. Our threat models for this segment cover the device, the DERS distribution pathway, the EHR↔pump bidirectional integration, and the central management server.

    We model the DERS library distribution channel as a primary attack surface - because it is. We model the EHR auto-program path with explicit attention to integrity (what stops a compromised gateway from sending an order with the right patient and the wrong dose?) and to operator override behavior. We model the central server as both a tenant-isolated cloud and a possible on-prem deployment with weaker controls. We model wireless network exposure (the pumps live on hospital Wi-Fi, often with poor segmentation). And we model the maintenance-mode / service-port path that's been the entry point in real incidents. Output is STRIDE-per-element traced to ISO 14971 hazards, with the DERS-library-tampering scenario called out as its own risk-control branch - because reviewers in this segment now expect to see it.

    Attack surface

    Layers we exercise in this engagement

    The infusion / drug delivery system, from the outermost cloud and clinician surfaces down to the device itself. Highlighted layers are exercised by this medical device threat modeling.

    1. 01Hospital network Tested
    2. 02Server / EHR integration Tested
    3. 03Pump-management system Tested
    4. 04Pump firmware Tested
    5. 05Drug library updates Tested

    Layers shown outermost (top) to innermost (bottom). Dashed rows are part of the surrounding system but out of scope for this view.

    How the engagement runs

    Medical Device Threat Modeling engagement, end to end

    Four phases, fixed fee, scoped to infusion / drug delivery architecture from kickoff onward.

    1. 01

      Design intake

      DFD3 data flow diagrams, trust boundaries, numbered entry points, and an Entry Point Descriptions table prepared from your design inputs.

    2. 02

      STRIDE analysis and attack trees

      Threat classification per in-scope element, with confidentiality, integrity, and availability attack trees. STRIDE classifies threats; it does not score them.

    3. 03

      Threat Table and traceability

      Threat IDs, initial risk, named controls, applicable test-case IDs, and manufacturer-supplied patient-harm context. Residual risk belongs in the separately scoped cybersecurity risk assessment.

    4. 04

      Threat-model handoff

      Threat model and Threat Table prepared for the applicable eSTAR attachments. The Global System, Multi-Patient Harm, Updatability and Patchability, and Security Use Case security architecture views are separate, separately scoped and priced deliverables. Testing and broader SPDF documentation are separate work items.

    Common findings

    What we see in Infusion / Drug Delivery medical device threat modeling

    The patterns we hit in this segment, this service, again and again.

    • DERS library distribution channel not threat-modeled

      Library treated as 'configuration data', not as code. Tampering, replay, and downgrade scenarios absent.

    • EHR auto-program integrity assumed

      Bidirectional integration trusts upstream order without per-message integrity. Compromised gateway scenario not modeled.

    • Maintenance / service port persistent across deployments

      Service interface threat-modeled in spec but enabled by default in the field. Discrepancy between as-designed and as-deployed not captured.

    • Wireless segmentation assumed

      Threat model presumes pumps on a dedicated VLAN. Don't assume the hospital segments its network. The threat model should assume mixed-VLAN exposure.

    Notable incidents

    Public infusion / drug delivery cybersecurity history

    Recalls, CISA ICS-MA advisories, and disclosed research that shape what reviewers ask about in this segment - and what our scope is built to cover.

    "Blue Goat's knowledge of regulatory requirements versus cybersecurity challenges was highly valuable and readily apparent as we were guided by and worked alongside their team towards the development of a comprehensive and compliant cybersecurity plan for our new medical device. Especially helpful for our company as we are a startup. Their team and competencies nicely filled our resource needs. Thank you Blue Goat!"
    Tim Luddy
    Tim Luddy
    Quality Manager · Retia Medical
    What you get

    Standard Medical Device Threat Modeling deliverables

    The same deliverables the parent Medical Device Threat Modeling service ships with - tuned to your infusion / drug delivery architecture.

    • ANSI/AAMI SW96 + ISO 14971 alignment
    • Full medical device system coverage
    • Threat-to-mitigation traceability
    • Justified methodology and assumptions
    Deliverable preview

    What lands in your eSTAR submission

    Documents structured for the cybersecurity attachments of your submission - no reformatting on your side.

    • ANSI/AAMI SW96 + ISO 14971 alignment
    • Full medical device system coverage
    • Threat-to-mitigation traceability
    • Justified methodology and assumptions
    Standards

    Standards that apply

    The Infusion / Drug Delivery baseline, plus the call-outs that matter for medical device threat modeling in this segment.

    FDA 2026 Premarket Cyber Guidance
    AAMI SW96
    IEC 62304
    ISO 14971
    IEC 81001-5-1

    Segment-specific call-outs

    IEC 60601-2-24 + AAMI TIR60601-4-5

    Pump-specific essential performance and security risk-control framing. Both are referenced.

    FDA infusion-pump-specific postmarket history

    Reviewers in this segment will compare your threat model against published incident patterns. Coverage gaps get flagged.

    Honest scoping

    What's not in scope

    We scope tightly on purpose. These items are either out-of-scope by design or belong in a separate engagement - we'll tell you up front, not after kickoff.

    • Penetration testing execution (scoped separately)
    • Clinical risk analysis under ISO 14971 (we feed it, we do not own it)
    • Hospital network architecture review
    FAQs

    Medical Device Threat Modeling for Infusion / Drug Delivery - FAQs

    The questions buyers in this segment actually ask before scoping a medical device threat modeling engagement.

    Related reading

    Go deeper on Infusion / Drug Delivery and premarket

    Pair this with

    Other engagements for Infusion / Drug Delivery

    Teams in this segment commonly bundle these alongside medical device threat modeling.

    Keep going

    Cybersecurity alerts and guides

    Goat's Weekly alerts

    Medical device advisories and safety communications, with links to original sources. An alert is a reason to assess applicability, not proof that your device is affected.

    Threat modeling, eSTAR 7.1, and SBOM evidence

    Connect threats to controls and test evidence. Compare the CISA 2026 SBOM fields with the NTIA 2021 baseline the FDA still cites, plus its support fields. Guidance recommendations remain separate from statutory duties.

    Threat modeling scope

    What is in a threat modeling engagement?

    Included work is defined in the signed proposal. These scope options are not automatic inclusions in every engagement.

    Fixed-fee deliverables

    • A threat model with DFD3 data flow diagrams, system elements, trust boundaries, and numbered entry points.
    • An Entry Point Descriptions table, STRIDE analysis per in-scope element, and confidentiality, integrity, and availability attack trees.
    • A Threat Table with Threat IDs, initial risk, named controls, and traceability to applicable test-case IDs and manufacturer-supplied patient-harm context.
    • Threat model and Threat Table documentation prepared for the applicable eSTAR cybersecurity attachments. Test execution and the residual cybersecurity risk assessment are separate deliverables.

    Required inputs

    • System architecture, interface list, intended use, and the device versions in scope. Include connected mobile apps, cloud APIs, update paths, and external dependencies where applicable.
    • Existing risk management file, manufacturer-defined harm severity and acceptance criteria, and any prior threat model.
    • Access to engineers who can answer design questions.

    Exclusions and separate work

    • Global System, Multi-Patient Harm, Updatability and Patchability, and Security Use Case architecture views are separately scoped and priced.
    • Residual cybersecurity risk assessment, penetration test execution, SBOM work, and broader SPDF documentation are separate work items.
    • General software verification and validation, clinical performance testing, and full QMS implementation or certification are outside Blue Goat's product cybersecurity scope.
    • The manufacturer retains engineering changes, clinical-risk evaluation, risk acceptance, release approval, and regulatory submission decisions. No regulatory outcome is promised.

    Pricing and scope changes

    Fixed fee, quoted after a scoping call. Scope depends on the number of interfaces and system elements. The Global System, Multi-Patient Harm, Updatability and Patchability, and Security Use Case security architecture views are not included in standalone threat modeling. They are scoped and priced separately. A broader premarket engagement can include both as distinct deliverables when the proposal specifies them. Risk acceptance remains a documented manufacturer decision. A written fixed-fee quote after scoping. The proposal names deliverables, versions, assumptions, required inputs, exclusions, and agreed dates. Additional work requires a separately agreed scope change. Missing inputs, access restrictions, manufacturer fixes, and review feedback can affect delivery.

    Medical Device Threat Modeling · Infusion / Drug Delivery

    Scope a Medical Device Threat Modeling engagement for your infusion / drug delivery program.

    A 30-minute discovery call with Christian Espinosa, Founder and CEO, and Kristy Kennedy, VP of Sales.